ISO2007MPEComputeType1ContinuousCornealIRUnweightedValue

>Psychtoolbox>PsychRadiometric>PsychISO2007MPE

[val_UWattsPerCm2,limit_UWattsPerCm2] = ISO2007MPEComputeType1ContinuousCornealIRUnweightedValue(S,radiance_WattsPerSrM2,stimulusDurationSecs,stimulusAreaDegrees2)

Compute the unweighted IR radiation for Type 1 instruments as given on page 8, Table 2,
5.4.1.4.

Input spectrum is radiance in units of Watts/[sr-m2-wlinterval].

Also return the exposure limit for this quantity.

See page 6 for a definition of a Type 1 instrument. As far as I can tell, the key
criterion is that it doesn’t put out more light than exceeds the Type 1 limits.

If the exposure time is longer than 2 hours the specified limits should be reduced by
1/exposureDuration in hours. This routine implements that adjustment for its returned
limit value. It does not implement a further reduction of of the limit (by a factor of 2)
for microscopes and endoilluminators.

****************************************************************************
IMPORTANT: Before using the ISO2007MPE routines, please see the notes on usage
and responsibility in PsychISO2007MPE/Contents.m (type “help PsychISO2007MPE
at the Matlab prompt.
****************************************************************************

6/25/13 dhb Wrote it.